Section 7-3 - Gathering Troubleshooting Data
7-3-5 Capturing Network Logs with Network Sniffer (cont’d)
By IP and port: Select Capture filter.
5.) Select New.
6.) Edit name for example DICOM port 104.
7.) Enter the string with the following syntax: port xxx and host yy.yy.yy.yy, where xxx is the port
number of “My Computer” in the scanner and yy.yy.yy.yy is the IP address of the device under test
(PACS,server, etc).
8.) Press
OK
.
9.) Select Start.
10.)Minimize network sniffer window and initiate communication to the DICOM device (eg. send images
to the storage device or query Worklist).
Figure 7-7 Sniffer Capture Filter IP and Port