Chapter 8 Replacement procedures
8 - 151
8-6-8-1
Alphanumeric Keyboard assembly removal
Table 8-101 Alphanumeric (A/N) Keyboard removal
Steps
Corresponding Graphic
1.
The Alphanumeric Keyboard
Assembly is installed in the Lower
Operator Panel.
Remove the two shift keys (cap lock
arrows), using the key removal tool
[the “MacGyver” keycap removal
tool will work as well, e.g., a
paperclip] to expose the screws that
secure the keyboard to the Lower
Operator Panel.
NOTE: If the keycap plunger does
not come off with the key, you have
to re-install it.
Keycap Tool
Keyboard Mounting Screw
NOTICE
To avoid damaging the keytops, use a small Phillips screwdriver to remove the keyboard
screws.
2.
Remove the Alphanumeric Keyboard screws (M3 x 8).