Section 8-7 - Main Console parts replacement
8-7-3-2 Removal Process with Non-loop Type Strap (cont’d)
8-7-3-2-1
Cleaning the bottom filter
The bottom filter is held in place with magnets. To clean the bottom air filter,
1.) You need to kneel down on the floor to access the bottom filter. Locate the piece of cloth extending
from the air filter on the bottom of the LOGIQ E9.
2.) You need two hands to remove this filter. One hand is used to pull on the piece of cloth that extends
from the filter and the other hand is used to wedge a finger just under a corner of the filter. In this
step, firmly take hold of the piece of cloth that extends from the filter.
3.) Position your other hand at a corner of the filter and wedge a finger between the filter and the bottom
of the LOGIQ E9.
4.) Pull down on the filter with the finger that is wedged between the filter and the LOGIQ E9 while
pulling on the piece of cloth that extends from the filter at the same time to remove the filter.
5.) Clean this filter with soap and water, or vacuum clean.
6.) Reposition the filter at the bottom of the LOGIQ E9.
Figure 8-54 Bottom View for Removing the Filter
WARNING
WARNING
Tilting the system is NOT a recommended method for the removal of the filter. The
bottom view in Figure 8-54 is only for illustrative purposes.