Chapter 8 Replacement procedures
8 - 183
8-6-14
Probe Cable Hooks replacement
8-6-14-1
Manpower
One person, 15 minutes.
8-6-14-2
Tools
8-2-5 "Tools needed for servicing the LOGIQ E9" on page 8-5
8-6-14-3
Probe Cable Hook removal
1.) Unscrew the fastening screw.
The screw may be locked with Lock-Tite, so you may need to apply a little extra force.
2.) Remove the Cable Hook.
8-6-14-4
Probe Cable Hook installation
1.) Position the Probe Cable Hook (Figure 8-45).
2.) Fix it in place by using an M5x20 screw with an M6 washer. Lock-Tite has to be used.
Torque: 50 Ncm (0.37 lbf-ft or 4.4 lbf-in) (fingertight).
Figure 8-45 Screw placement, Probe Cable Hook