Chapter 7 Diagnostics/Troubleshooting
7 - 69
7-5-9-10
Network Adapter
7-5-3 "Diagnostics Window Overview" on page 7-16
for a description of this window’s features.
Runs a test on essential functions of the Network Interface. The test requires the system to be
connected to a network. The test results are displayed in the Status portion of the window.
This test gives the current status of the network and provides the option to restart the network
components in the BEP.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
Click the Execute button.
7-5-9-11
Network Adapter Diagnostic Changes for R4
Network Adapter Test is split into 2 folders:
Each folder contains separate tests for Wired and Wireless Adapters.
7-5-9-12
Memory
7-5-3 "Diagnostics Window Overview" on page 7-16
for a description of this window’s features.
Runs a test on essential functions of the Memory. The test results are displayed in the Status portion of
the window.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Replace the BEP if any tests fail.