GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 99
4.
Lay the Monitor flat.
Remove the two lower Phillips screws (
1
).
Loosen the upper Phillips screws (
2
). The
Monitor will be supported by the two upper
screws.
Slide the Monitor off of the Arm Assembly.
Remove the upper Phillips screws to re-
install into the replacement Monitor.
19 and 23 inch Monitor (19 inch shown)
Upper Monitor mounting screw (2) and hole
OLED Monitor mounting screws and holes
Table 8-64 Monitor removal - R5.x and later (Ergotron Arm)