![Dentsply Sirona Orthophos SL 2D Operating Instructions Manual Download Page 97](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684097.webp)
64 94 954 D3632
D3632
.
201.03.08
.
23 07.2018
97
Dentsply Sirona
5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.4.7.1
Positioning with 3D bite block or universal bite block
In general, exposures of all volume areas can be made with the 3D bite
block, the yellow bite block or the universal bite block. The 3D bite block
or the universal bite block is recommended for volume exposures, as the
patient is positioned more securely. If it is not possible to work with these
bite blocks, another possibility is to use the chin rest with the bite block
rod or bar, see "Positioning with chin rest [ → 99]". The foam of the
universal bite block is equally suitable for patients without anterior teeth.
✔
The 3d bite block, the yellow bite block, the universal bite block and
forehead support and temple supports are inserted in the unit.
✔
The relevant hygienic protective sleeves are pulled over the
accessories.
1. Guide the patient in front of the control mirror.
2. Set the unit height using the up A and down B keys.
CAUTION! The height adjustment motor starts slowly and then
increases its speed.
Press and hold down the button until the desired height is reached.
The unit movement is accompanied by a beep.
Release the height adjustment buttons once the patient's mouth and
the bite block are at the same height.
3. Instruct the patient to hold the handles with both hands and bite into
the bite block.
ª
The patient's anterior teeth are positioned in the indentation in the
bite block. If necessary, push the lower anterior teeth forwards
until they reach a stop.
4. Swivel the control mirror outwards. Press the left recess on the
touchbar D.
ª
You can see the patient in the control mirror.
B
A
Summary of Contents for Orthophos SL 2D