![Dentsply Sirona Orthophos SL 2D Operating Instructions Manual Download Page 144](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684144.webp)
64 94 954 D3632
144
D3632
.
201.03.08
.
23 07.2018
7 Malfunctions
Dentsply Sirona
7.1 Help messages
Operating Instructions Orthophos SL 2D / 3D
H322 – Select quadrant
No quadrant is selected.
➢ Select the desired quadrant.
ª
The program sequence is continued.
H325 – Select region of exposure
No volume area selected.
➢ A dental arch showing the volume areas is displayed on the
touchscreen. Touch an area to select it.
ª
The program sequence is continued.
H403 – Switch Sidexis to ready for exposure state
Sidexis is not ready for exposure.
➢ Switch Sidexis to a ready-for-exposure state; see the Sidexis user
manual.
H406 – R button, move into Ceph starting position
Ceph is not in the starting position.
➢ Press the R key.
ª
The unit travels to the starting position.
H420 – Rescue: do not switch off; see Rescue program for issues
during image transfer and raw data import [ → 121]
The image could not be transferred to Sidexis.
Switching the unit off, SiRescue
➢ Save the exposure using Unit Rescue.
ª
The image is transferred to Sidexis.
CAUTION
The exposure memory of the unit is cleared when the unit is switched
off.
Images that have not been transferred to Sidexis cannot be restored
using Unit Rescue.
➢ Never switch off the unit before transferring the images to Sidexis.