64 94 954 D3632
56
D3632
.
201.03.08
.
23 07.2018
5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.1.2
Preparing the exposure
Depending on the patient or the exposure program, accessory parts must
be replaced and the appropriate exposure mode must be selected; see
"Installation and start-up" [ → 43].
You will require the following accessories:
● Chin rest with bite block rod or bar or
yellow bite block or contact segment or
universal bite block or occlusal bite block.
● Temple supports
● Forehead support
➢ Insert the accessory parts to be used into the unit and pull on the
relevant hygienic protective sleeves, see "Hygienic protective
➢ Switch Sidexis to a ready-for-exposure state; see ‘Switch Sidexis to
ready-for-exposure state’.
CAUTION
The chin rest must not be used for children when using programs BW1
and BW2. The positioning is otherwise too low.
IMPORTANT
For programs BW1 and BW2, the universal bite block must not be used
above the black marking. The positioning is otherwise too low.