![Dentsply Sirona Orthophos SL 2D Operating Instructions Manual Download Page 93](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684093.webp)
64 94 954 D3632
D3632
.
201.03.08
.
23 07.2018
93
Dentsply Sirona
5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.4.4
Adjusting the volume area and collimation
You can change between VOL1, VOL2, and VOL3 (optionally) depending
on whether a smaller or larger volume can be used.
For volume exposures one of the predefined volume areas must be
selected. By selecting the quadrants, collimation of the volume in VOL1
program can be set to a height of 5.5 cm for the upper or lower jaw area,
while in VOL2 program, the height is permanently set to 5.5 cm and in
VOL3 program, collimation of the volume can be set to a height of 7.5 cm
for the upper jaw and 8.0 cm for the lower jaw. However, you must always
select a quadrant; or, for the front upper jaw teeth or front lower jaw teeth,
the upper or lower area must be selected.
✔
Level 1 is displayed on the touchscreen.
✔
The message "H403 – Switch Sidexis to ready for exposure state"
appears in the comment line.
1. Select the desired volume area A. Touch one of the circles for the
anterior, molar, or temporomandibular joint area (only VOL1, VOL3)
in the center of the touchscreen.
ª
The selected volume area is highlighted orange. According to
your selection, the quadrant selection presetting B changes if
necessary:
VOL1, VOL3: The quadrant selection can be used for volume
collimation. Selecting the molar or temporomandibular joint
region, for example, sets collimation for the upper or lower jaw
region.
2. Press the quadrant symbol (B) on the right side of the touchscreen.
ª
A submenu line is opened.
IMPORTANT
The regions shown do not correspond to the actual diameter of the
volume on the touchscreen. See the figure in the section "VOL1 –
Program description".
Summary of Contents for Orthophos SL 2D
Page 2: ......