3 Unit description
Dentsply Sirona
3.2 Technical data
Operating Instructions Orthophos SL 2D / 3D
3.2
Technical data
3.2.1
Unit data
Model designation:
Orthophos SL
Nominal voltage:
200 – 240 V
Permissible fluctuation:
± 10%
Permissible drop under load: 10%
Rated current:
12A
Nominal power output:
2 kW at 90 kV/12 mA with any radiation
time
Nominal frequency:
50 Hz / 60 Hz
Mains resistance:
max. 0.8 ohms
Main building fuse:
25 A slow-blow (16 A for single line)
Power consumption:
2 kVA
Power output of tube
assembly:
69 kV / 16 mA = 1104 W with any
radiation time
Tube voltage:
60 – 90 kV (for 90 kV max. 12 mA)
Tube current:
3 – 16 mA (for 16 mA max. 69 kV)
Maximum setting range:
60 kV / 3 mA to 90 kV / 12 mA
High-voltage waveform:
High-frequency multipulse
Residual ripple ≤ 4 kV
High voltage generation
frequency:
40 – 120 kHz
Program duration:
See "Program values" [ → 148].
Exposure time:
See "Program values" [ → 148].
Image acquisition scale:
For P1, normal dental arch (slice center)
approx. 1:1.19, i.e. the acquired image is
magnified by approx. 19% on average
compared to reality.
Exposure time for a
cephalometric image:
14.9 s max.
Image acquisition scale for a
cephalometric image:
approx. 1:1.1, i.e. the acquired image is
magnified by approx. 10% on average
compared to reality.
Total filtration of X-ray tube
assembly:
> 2.5 mm AI / 90 IEC 60522
0.3 mm Cu for volume exposures
1 mm Cu for volume exposures in Low
Dose mode
Focal spot size as specified
in IEC 60336, measured in
the central X-ray beam:
0.5 mm