2 Safety information
Operating Instructions Orthophos SL 2D / 3D
2.5 Qualifications of operating personnel
2.5
Qualifications of operating personnel
The unit may only be operated by skilled or properly trained personnel.
Personnel, who are to be trained, taught, instructed or are taking part in
a general training, may operate the device only under the supervision of
an experienced person.
To operate the unit, the operating personnel must:
● have read and understood the Operating Instructions
● be familiar with the fundamental structure and functions of the unit
● be able to recognize irregularities in the functioning of the unit and
implement the appropriate measures where necessary
The operators of the unit must be adequately trained in the DVT
technology with regard to the principles of operation and radiation
protection.
2.6
Switch on the unit
Safety information for switching on the unit: Customer
The patient may not be positioned in the unit when starting it and setting
the operating mode (until the sensor has been positioned).
Malfunction may cause injury to the patient.
In case of an error that requires switching the unit off and back on again,
the patient must be removed from the unit, at the latest before the unit is
switched back on.
2.7
Radiation protection
The valid radiation protection regulations and measures must be
observed. The statutory radiation protection equipment must be used. In
order to reduce radiation exposure, Dentsply Sirona recommends using
bismuth or lead shields or aprons, especially for pediatric patients.
During an exposure, the operator should move as far away from the X-ray
tube assembly as the coiled cable of the release button permits.
With the exception of the patient, no other persons without radiation
protection are allowed to stay in the room during an exposure. In
exceptional cases, a third person may provide assistance, but not the
practice staff. Visual contact with the patient and the unit must be
maintained throughout the entire exposure.
In case of malfunctions, cancel the exposure immediately by letting go of
the exposure release button.
CAUTION
CAUTION!
Not for use on patients less than approximately 5 years old [~21 kg (46
lb) and 113 cm (44.5 in) standing height]