5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.5.3
Selecting an exposure program
✔
The unit is switched on and ready for exposure.
1. Touch the CEPH symbol at the top of the touchscreen.
ª
The CEPH program group is selected.
2. Select an exposure program. To do this, touch the arrow keys + C
and - A. If you want to select a subprogram, such as C3F, touch the
program display B repeatedly. All the subprograms in the selected
program are displayed one after the other.
3. Follow the instructions in the comment line on the touchscreen. Press
the R key if necessary.
ª
The diaphragm and the sensor move into the starting position for
cephalometric exposures.
ª
The exposure program is selected.
5.1.2.5.4
Adjusting the collimation
In programs C3, C3 F, C1 p.a. and C2 a.p, the exposure area can be
limited so that certain parts of the patient's head, e.g. the upper cranial
region, thyroid gland and posterior cranial region, are not irradiated. This
reduces the patient dose.
✔
Level 1 is displayed on the touchscreen.
1. Press the collimation symbol (A) on the right side of the touchscreen.
ª
A submenu line is opened.
2. Select the collimation.
ª
The selection is highlighted orange in the submenu line.
3. Touch the cross on the left side of the submenu line.
or
➢
Press the collimation symbol (A) again.
ª
The submenu line is closed.
ª
The collimation is set.
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
C
B
A
Ready for exposure
A
Ready for exposure
Ready for exposure