5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.5.2
Preparing the exposure
Depending on the patient or the exposure program, accessory parts must
be replaced and the appropriate exposure mode must be selected; see
"Installation and start-up" [ → 43].
The following illustrations of the cephalometer are shown in the left-
handed arm version. They also apply for the cephalometer with a right-
handed arm.
A = asymmetrical
S = symmetrical
Preparing the nose support
1. Grasp the nose support at the rotary joint.
2. Pull the nose support forwards until it reaches the stop.
3. Swivel the nose support sideways and upwards.
Move the ear plug holders
1. Grasp the holders at the very top with both hands.
2. Push the holders simultaneously outwards as far as they will go.
NOTE
The adjustment of the cephalometer may alter depending on the load.
A change in the adjustment may lead to faulty X-rays.
➢ Never lean against the cephalometer or the extension arm.
➢ Do not hang or place any objects against or on the cephalometer or
extension arm.