![Dentsply Sirona Orthophos SL 2D Operating Instructions Manual Download Page 64](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684064.webp)
64 94 954 D3632
64
D3632
.
201.03.08
.
23 07.2018
5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
9. Align the patient with the central laser line G.
ª
The laser line reflects in the center of the patient’s anterior teeth
or the middle of the face (mid-sagittal).
10. If necessary, check the patient’s position. Briefly press the up A and
down B height adjustment keys.
11. Press the temple support adjustment button I.
ª
When you touch the patient’s head, the temple supports stop
automatically.
12. Press the forehead support adjustment key (H).
ª
When you touch the patient’s head, the forehead support stops
automatically.
Make sure that the patient’s head does not move backwards
when placing the forehead support.
13. Instruct the patient to exhale, place his tongue against the roof of his
mouth, and hold this position until the end of the exposure.
ª
The patient is positioned in the unit.
If the occlusal bite block is stuck in the bite block holder even after the
exposure is carried out and you select an exposure program that is not
intended for the use of the occlusal bite block, the help message ‘H307 -
Change bite block ’ will appear in the comment line. Insert the relevant
bite block or contact segment required for the exposure. The help
message disappears as soon as the occlusal bite block is removed.
I
H
IMPORTANT
In P1, P2, P10, the slice width is selected automatically for different
dental arches with the temple support setting, and the radiation time is
also changed slightly through this in accordance with the temple support
width which is set.