7 Malfunctions
Dentsply Sirona
7.3 Error description
Operating Instructions Orthophos SL 2D / 3D
7.3
Error description
7.3.1
Ex – Error type
E1 – System warning/message
The error is in an acceptable tolerance range. Device operation is not
directly impaired.
1. Acknowledge the error message.
2. Contact your Customer Service.
ª
Continued device operation is ensured.
E2 – Overload
The error can be traced back to temporary overheating or something
similar.
1. Acknowledge the error message.
2. Wait for a moment and repeat the procedure step. If the error
reappears, extend the waiting time.
ª
The error no longer occurs after a certain waiting period.
3. If the error persists, contact your Customer Service.
E3 – Key pressed during power-up
The error results from an invalid signal state due to pushing buttons and
security signals during power-up.
1. Switch the unit off and on again. NOTE! Observe waiting period!
2. If the error persists, contact your Customer Service.
E4 – mechanical blocking
Errors that indicate mechanical blocking of motor-driven parts.
1. Check whether the device is mechanically blocked.
Remove items.
2. Switch on/off. Check whether the error reoccurs.
If the error remains permanently, inform customer services.
E5 – Malfunction during exposure or exposure preparation
Error resulting from a certain system action triggered by the user which
could not be performed because a required (internal) partial function
(software or hardware) is not ready or fails.
1. Acknowledge the error message.
2. Repeat the last procedure step or exposure.
ª
The error no longer occurs.
NOTE
The unit must not be switched on/off constantly.
Constant switching on and off reduces the service life of individual unit
components and results in increased power consumption.
➢ After switching the unit off, wait for approx. 60 seconds before
switching it on again.