![Dentsply Sirona Orthophos SL 2D Operating Instructions Manual Download Page 75](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684075.webp)
64 94 954 D3632
D3632
.
201.03.08
.
23 07.2018
75
Dentsply Sirona
5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
7. Align the patient with the central laser line (G).
ª
The laser line reflects in the center of the patient’s anterior teeth
or the middle of the face (mid-sagittal).
8. Align the patient's head according to the Frankfurt horizontal plane
(E).
9. Adjust the height of the light localizer using the slider (F).
ª
The laser line reflects on the upper edge of the outer ear canal.
10. Correct the patient's head inclination as necessary. Briefly press the
up (A) and down (B) height adjustment keys.
ª
The laser line reflects on the lowest point of the lower eye socket
edge.
11. Press the forehead support adjustment key (H).
ª
The forehead supports stop moving automatically when they
come into contact with the patient's forehead.
Ensure that the patient's head does not move backward when the
forehead support is put in place.
12. Check the patient's position and make any final corrections as
necessary.
13. Swivel the control mirror back into place. by pressing the right recess
on the touchbar (D).
ª
The patient can see himself in the control mirror.
14. Instruct the patient to exhale, place his tongue against the roof of his
mouth, and hold this position until the end of the exposure.
ª
The patient is positioned in the unit.
E
B
A
1
H
IMPORTANT
Note the information on the two-part exposure programs in the
"Releasing the exposure" chapter, see “Two-part exposure programs