6 Service
Operating Instructions Orthophos SL 2D / 3D
6.1 Cleaning and care
6.1.3
Sterilization
The following accessories can be sterilized:
In addition, also use the hygienic protective sleeves; see ‘Hygienic
WARNING
Infections can be transmitted from patient to patient.
Accessories that are not sterilized correctly can cause illness in patients.
➢ All accessories that are suitable for sterilization should only be
sterilized in an autoclave at 132 °C (270 °F), with at least 4 minutes
holding time and at 2.1 bar (30.5 psi) overpressure.
WARNING
The hygienic protective sleeves are single use devices.
Contaminated hygienic protective sleeves can cause illness in patients.
➢ Replace the hygienic protective sleeves after each patient.