64 94 954 D3632
28
D3632
.
201.03.08
.
23 07.2018
3 Unit description
Dentsply Sirona
3.3 Overview of exposure programs
Operating Instructions Orthophos SL 2D / 3D
Volume exposure
The Orthophos SL X-ray system is available as a 2D or 2D/3D hybrid
device. 2D devices are prepared for subsequent upgrades with a 3D flat
panel detector. The volume programs VOL1 SD, VOL1 HD, VOL1 Low,
VOL2 SD, VOL2 HD, VOL2 Low (VOL3 SD, VOL3 HD and VOL3 Low as
an option) are only available in addition with this.
The Low Dose mode in programs VOL1, VOL2 und VOL3 is not available
in every country.
Program settings: Volume area (anterior teeth, molars right/left or
temporomandibular joints right/left), collimation of upper/lower jaw,
radiation time
For more information about the 3D exposure program, see "Volume
Programs
Volume area
Collimation
VOL1 SD
Isotropic voxel edge length: 160 µm
VOL1 HD
Isotropic voxel edge length: 160 µm
VOL1 Low
Isotropic voxel edge length: 160 µm
Volume exposure with a diameter of approx. 8 cm and a
height of approx. 8 cm or 5.5 cm collimated.
VOL2 SD
Isotropic voxel edge length: 160 µm
VOL2 HD
Isotropic voxel edge length: 80 µm
VOL2 Low
Isotropic voxel edge length: 160 µm
Volume exposure with a diameter of about 5 cm and a
height of about 5.5 cm for upper or lower mandible
Optional
VOL3 SD
Isotropic voxel edge length: 220 µm
VOL3 HD
Isotropic voxel edge length: 160 µm
VOL3 Low
Isotropic voxel edge length: 220 µm
Volume exposure with a diameter of about 11 cm and a
height of about 10 cm or selection of upper quadrant
collimated to 7.5 cm and selection lower quadrant
collimated to 8.0 cm
Summary of Contents for Orthophos SL 2D