![Dentsply Sirona Orthophos SL 2D Operating Instructions Manual Download Page 150](http://html1.mh-extra.com/html/dentsply-sirona/orthophos-sl-2d/orthophos-sl-2d_operating-instructions-manual_2483684150.webp)
64 94 954 D3632
150
D3632
.
201.03.08
.
23 07.2018
8 Settings and repair
Dentsply Sirona
8.1 Program values
Operating Instructions Orthophos SL 2D / 3D
8.1.2
Panoramic exposure index 2E
Code 2E
It guarantees that the applicable legal regulations which must be
complied with since January 1, 1999 are strictly observed. Furthermore,
this level series can also be applied worldwide. National regulations must
be complied with. The table lists the maximum exposure times.
Level series for code 2E
Program
Pro
-
gram
duration
approx.
Max. ex
-
posure
time
Quickshot
program
duration
approx.
Max.
Quickshot
exposure
time
Factory setting
User-defined values
– Please enter here –
P1
19.0s
12.9s
14.1s
8.0s
14.2s
10.3s
9.0s
5.1s
63/6 63/8 69/8 72/8
P1A
21.8s
15.4s
14.1s
8.0s
18.2s
13.9s
9.0s
5.1s
63/6 63/8 69/8 72/8
P1C
20.1s
13.3s
14.1s
8.0s
17.1s
12.6s
10.5s
5.9s
63/6 63/8 69/8 72/8
P2
16.4s
11.6s
11.5s
6.7s
12.4s
9.4s
7.3s
4.2s
63/6 63/8 69/8 72/8
P2A
18.0s
12.1s
11.5s
6.7s
15.0s
11.8s
7.3s
4.2s
63/6 63/8 69/8 72/8
P2C
16.8s
11.7s
11.5s
6.7s
13.7s
9.7s
8.5s
4.9s
63/6 63/8 69/8 72/8
P10
16.4s
11.6s
11.5s
6.7s
11.4s
9.4s
7.3s
4.2s
63/6 63/8 69/8 72/8
P10A
18.0s
12.1s
11.5s
6.7s
15.0s
11.8s
7.3s
4.2s
63/6 63/8 69/8 72/8
P10C
16.8s
11.7s
11.5s
6.7s
13.7s
9.8s
8.5s
4.9s
63/6 63/8 69/8 72/8
P12
11.9 s
4.9 s
69/8 75/8 78/7 84/6
BW1
23.0s
23.0s
8.8s
4.5s
63/6 63/8 69/8 72/8
BW2
18.0 s
5.1 s
63/6 66/8 69/8 72/8
TM1.1+TM1.2 16,1+
16.1s
6,4+
6.4s
66/8 69/8 72/8 75/8
TM3
18.4 s
8.1 s
63/8 66/8 69/8 72/8
S1
19.8 s
14.4 s
69/8 75/8 78/7 84/6
S3
20.0 s
8.1 s
69/8 75/8 78/7 84/6
Summary of Contents for Orthophos SL 2D