141
Dentsply Sirona
6 Service
Operating Instructions Orthophos SL 2D / 3D
6.2 Inspection and maintenance
6.2
Inspection and maintenance
Inspection and preventive maintenance must be performed at scheduled
intervals to protect the health and safety of patients, users, and other
persons.
The information provided in the document
"Inspection and maintenance
and safety-related checks"
REF
"64 95 100"
should be helpful here. The
document can be downloaded at http://www.dentsplysirona.com/
manuals.
Annual inspection
In order to ensure the operational safety and functional reliability of your
product, you as the system owner should check the equipment at regular
intervals (at least once a year) or commission your dental depot to do so.
Maintenance by the service engineer
In addition to the annual check to be carried out by the system owner or
authorized persons, preventive maintenance must be performed after 4,
7 and 10 years, and then at two-year intervals.
Checking image quality
The image quality should be assessed by the system owner at regular
intervals, at least once a year.
On digital image receptor systems, the degree of postprocessing
(brightness or contrast adjustment) that is required in the image
processing software (e.g. Sidexis) to produce satisfactory results is used
as an assessment criterion.
Checking image quality
Constancy test in Germany
Unit:
Please note the legally required regular checks of the image quality of
your X-ray device as stipulated in your country (e.g., RöV, the German X-
ray Ordinance). Dentsply Sirona provides you with the Sidexis software
for easy implementation of this constancy test and for its documentation.
The required test specimen and the description of the constancy test
come with the unit.
Diagnostic monitor:
In Germany, a regular constancy test of the results monitor is required by
the X-ray Ordinance in accordance with DIN 6868-157. To easily
implement the necessary legal requirements, Dentsply Sirona provides
the Simocon software. You can find this software and the operating
instructions that belong to it on your Sidexis CD under the "Tools" section.