5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.2 Preselecting basic settings
Start settings: Changing the start settings for the patient symbol and
selecting enabling/disabling the Quickshot function
✔
Level 2 of the CEP program group is displayed on the touchscreen.
1. Touch the diskette symbol (C) at the top of the touchscreen.
ª
The start settings are displayed.
2. Select the patient symbol that you want to preset.
ª
The selection is highlighted in orange and displayed on the right
side of the touchscreen.
3. Select whether the Quickshot function should be switched on or off.
Press the
Quick On
or
Quick Off
icons on the touchscreen. The
Quickshot function reduces the cycle time by approx. 20 to 50%,
depending on the exposure program. The function is set
independently of the exposure program.
ª
The selection is highlighted in orange and displayed on the right
side of the touchscreen.
4. Touch the save symbol (B).
ª
The setting is saved for the CEPH program group.
5. Touch the double triangle D in the upper right corner of the
touchscreen.
ª
The start settings are hidden, level 1 is displayed.
C
Select basic settings
B
D
Select start settings