64 94 954 D3632
126
D3632
.
201.03.08
.
23 07.2018
5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.3.4
Using the remote control
On the remote control, exposures are triggered using the release button
(D). If it is not possible to maintain visual contact with the patient when
releasing the exposure, the release key with the coiled cable (F) on the
X-ray unit can be removed and used on the remote control.
If the unit is ready for exposure and no auxiliary messages are displayed,
the current program parameters appear on display (C): Program
designation, exposure time, voltage, current in the individual fields (
Prog.
,
s
,
kV
,
mA
). The exposure can be released now.
If plain-text help messages are displayed on the Easypad touchscreen,
they also appear in coded form on the
Prog.
display of the remote control,
alternating with the program name.
When switching on the unit, the X-ray indicator A lights up for a functional
check for approx. 1 second.
The LED B lights up when the unit is on.
Use the Return key E to acknowledge exposures, error messages and
auxiliary messages and move the rotating unit to the starting position.
If a row of dots appears in the
Prog.
field on the digital display (C), this
means the unit is currently in a preparatory phase (e.g. unit movements,
parameter settings, program loading times, etc.). Wait until the dots
disappear automatically and the system is ready for operation.