7 Malfunctions
Operating Instructions Orthophos SL 2D / 3D
7.1 Help messages
7
Malfunctions
7.1
Help messages
When working with the unit, auxiliary messages are displayed for certain
actions (e.g. press H301 for the return key), which call for the user to
perform a specific action. These auxiliary messages are listed below. If
an error occurs, error messages are output starting with "E" followed by
5 digits, see "Error description" [ → 146].
✔
The unit is switched on and ready for operation.
1. Press the release button.
ª
The message H3/H4 xx appears.
2. See list below about how to proceed to make the system ready for
exposure.
H301 – R button, move into starting position
The rotating unit is not in the starting position.
XGPF-63464; Moving into the starting position
➢ Press the R key.
ª
The unit travels to the starting position.
H307– Change bite block
The occlusal bite block cannot be used for the selected exposure
program.
➢ Remove the occlusal bite block from the unit, use the positioning aid
that matches the exposure type.
ª
The program sequence is continued.
H320 – R button, confirm exposure data
The exposure data have not been acknowledged yet.
➢ Press the R key.
ª
Exposure data are confirmed.
H321 – Close the door
Check door contact of the X-ray room.
➢ Close the door to the X-ray room.
ª
The contact switch on the door is closed.
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.