5 Operation
Dentsply Sirona
5.2 Preselecting basic settings
Operating Instructions Orthophos SL 2D / 3D
5.2.1.3
CEPH program group
In level 2:
In the basic settings, the stored kV/mA values for the patient symbols can
be adjusted for each program.
In the default settings, the pre-selection of the patient symbol can be
changed and the enabling/disabling the Quickshot function can be preset.
Basic settings: changing the kV/mA values of the patient symbols
✔
Level 1 of the CEPH program group is displayed on the touchscreen.
1. Touch the toothed wheel A in the upper right corner of the
touchscreen.
ª
Level 2 is displayed.
2. Select the exposure program for which you want to change the kV/
mA value. Touch the - or + arrows.
ª
The selected setting is displayed on the right side of the
touchscreen.
3. Select the patient symbol for which you want to change the kV/mA
value.
4. Select the kV/mA value that you want to apply to the selected
program and patient symbol.
5. Touch the save symbol (B).
ª
The kV/mA value is saved for the selected program and patient
symbol.
6. Repeat this procedure for the other patient symbols and programs.
ª
The settings for the kV/mA values stored for the patient symbols
have been made.
A
Ready for exposure
B
Select basic settings