5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5. Press return key (R) on the Easypad.
ª
Confirmation of the exposure data is acknowledged.
6. WARNING! The patient may be injured by moving parts. Press return
key R on the Easypad again.
ª
The rotating unit moves to the starting position.
ª
The unit is ready for the next exposure.
Save data; see Sidexis
Two-part temporomandibular joint exposure program TM 1.1
In the Two-part exposure program TM 1.1, two exposures (TM 1.1 and
TM 1.2) are made.
✔
Once the first temporomandibular joint exposure has been released,
the message
"Please wait"
appears in the comment line on the
touchscreen.
✔
The rotating unit has automatically moved to the starting position.
1. Instruct the patient to open his/her mouth.
ª
The patient has opened his/her mouth without changing his/her
position.
2. Press the release button A and hold it down until the end of the
second exposure.
ª
The second exposure is released.
"Exposure is performed"
appears in the comment line on the touchscreen.
3. Wait until a short pulsed tone sequence sounds (can be deactivated
by a service engineer).
ª
The message
"Please wait"
appears in the comment line of the
touchscreen followed by confirmation of the exposure data.
Exposure mode, exposure program, tube voltage and current,
actual radiation time, dose-area product are displayed.
4. Let go of release button (A).
ª
The second exposure is completed. Proceed as described above
starting with step 5.
ª
The X-ray image is displayed on the PC monitor after a brief
period.
R
Ready for exposure