5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.3.3
Rescue program for issues during image transfer and raw data
import
In case of unexpected network errors or exposure interruptions, there
may be problems with image transfer to Sidexis 4.
There are two options available in Sidexis 4 for transferring images, which
are described below:
● The exposure data can still be found in the exposure memory of the
unit and auxiliary message H420 is displayed on the Easypad; see
● The exposure data has been transferred from the X-ray unit, but has
not yet been imported into Sidexis 4, see “Data container rescue
5.1.3.3.1
Unit rescue
✔
The auxiliary message H420 is displayed on the Easypad.
✔
The X-ray exposure could not be transferred, as a network fault exists
or the PC failed.
In this case, the system terminates the connection after a unit-
dependent time and switches to Rescue state.
This means: That the image is not lost, but is kept in the RAM of the
X-ray component by a safety mechanism until it is retrieved by
Sidexis. A further exposure with this X-ray unit is blocked until this
moment.
1. Select the X-ray unit in Rescue state.
2. Select the current exposure.
CAUTION
The exposure memory of the unit is cleared when the unit is switched
off.
Images that have not been transferred to Sidexis are irretrievably lost.
The exposure must be repeated.
➢ Wait until the exposure data have be completely transferred. Do not
switch the unit off before the X-ray exposure is displayed on the
Sidexis screen.