5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.1.4
Setting quadrants
The exposure can be restricted to quadrants. You can select a half-view
of the right or left jaw in programs P1, P2, P10 and BW1, or upper jaw/
lower jaw in programs P1, P2, P10 and P12. In P1, P2, P10, also for
constant magnification and artifact-reduced view.
✔
Level 1 is displayed on the touchscreen.
1. Press the quadrant symbol (A) on the right side of the touchscreen.
ª
A submenu line is opened.
2. Select the desired quadrant(s). See the figure below. Quadrants can
be selected as half-views or individually. Touch the quadrant field in
the center to reactivate the full frame.
ª
The selected quadrants are emphasized / the non-selected ones
are grayed out.
3. Touch the cross on the left side of the submenu line.
or
➢
Touch the quadrant symbol (A) again.
ª
The submenu line is closed.
ª
The quadrant or quadrants are set.
A
Ready for exposure
IMPORTANT
The program duration for individual quadrant exposures equals the
program duration for half-view exposures.
Ready for exposure
Ready for exposure
Ready for exposure
Ready for exposure
R
L
R
L
R
L
R
L