8 Settings and repair
Operating Instructions Orthophos SL 2D / 3D
8.1 Program values
Sample calculation
X-ray with program P1 and a kV/mA parameter combination of 60 kV/
10 mA
Ad 1. 60 kV has a DAP factor of 0.5693 at aperture 10
Ad 2. 10 mA displayed
Ad 3. the exposure time is 14.1 s
2D-images
3D-images
kV
DAP factor
program
P1/P2/P12/
S1/S3/TM1/TM3
(mGy x cm
2
/mAs)
DAP factor
program
P10/BW1/BW2
(mGy x cm
2
/mAs)
DAP factor
program
C1-C4
(mGy x cm
2
/mAs)
60
0.5693
0.3448
0.1024
62
-
-
0.1101
63
0.6308
0.3820
-
64
-
-
0.1178
66
0.6983
0.4229
0.1262
69
0.7570
0.4585
0.1387
71
-
-
0.1470
72
0.8305
0.5030
-
73
-
-
0.1556
75
0.8981
0.5439
-
77
-
-
0.1737
78
0.9679
0.5862
-
80
-
-
0.1868
81
1.0420
0.6311
-
84
1.1024
0.6677
0.2055
90
1.2360
0.7486
0.2345
DFP
0 5693
mGycm
2
mAs
----------------------
,
10
mA
×
14 1
s
,
×
80 2713
mGycm
2
,
=
=
kV
3D
exposure
mode
DAP factor
program
VOL1
(mGy x cm
2
/mAs)
DAP factor
program
VOL2
(mGy x cm
2
/mAs)
DAP factor
program
VOL3
(mGy x cm
2
/mAs)
8x8
8x5.5
5x5.5
11x10
11x7.5/8
85
SD
9.459
6.583
4.191
15.401
11.790
85
HD
11.094
7.722
4.951
18.064
13.829
85
Low Dose
3.059
2.129
1.365
4.981
3.813