5 Operation
Dentsply Sirona
5.1 Acquiring the X-ray image
Operating Instructions Orthophos SL 2D / 3D
5.1.2.2.5
Setting the kV/mA values
Setting the kV/mA values via the patient symbols
Preset kV/mA combinations are assigned to the patient symbols, which
are selected according to the patient's size and weight. The symbols
roughly correspond to child, youth/woman, woman/man, hefty persons.
✔
Level 1 is displayed on the touchscreen.
➢ Touch the required patient symbol.
ª
The selection (A) is highlighted in orange. The selected kV/mA
value is displayed on the right side of the touchscreen.
ª
The kV/mA value is set.
For the USA only: The source to image distance (SID) is displayed on the
touchscreen.
This is 497 mm (19.6“) for the PAN sensor and
524 mm (20.6“) for the Flat Panel Detector.
Setting the kV/mA values via the submenu line
If the preset kV/mA combinations do not provide a satisfactory result, you
can also set the kV/mA values manually in all programs.
✔
Level 1 is displayed on the touchscreen.
1. Touch the kV/mA symbol (B) on the right side of the touchscreen.
ª
A submenu line is opened.
2. Select a kV/mA value. Touch the - or + keys.
ª
The selected kV/mA value is displayed.
3. Touch the cross on the left side of the submenu line.
or
➢
Touch the kV/mA symbol (B) again.
ª
The submenu line is closed.
ª
The kV/mA value is set.
Ready for exposure
A
B
Ready for exposure
Ready for exposure
Ready for exposure