4 Installation and start-up
Dentsply Sirona
4.3 Removing/inserting the Ceph sensor
Operating Instructions Orthophos SL 2D / 3D
4.3
Removing/inserting the Ceph sensor
The Ceph sensor must always be inserted in order for the unit to be
operated. However, should the Ceph sensor need to be removed, please
proceed as follows:
Removing the sensor
1. Hold the sensor firmly.
2. Press the button all the way in and hold it.
ª
The sensor is released from the fastening.
3. Pull the sensor downwards out of the guide.
Inserting the sensor
1. Hold the sensor firmly.
2. Using both guide pins, insert the sensor into the guide sleeves on the
unit and push until it reaches a stop.
ª
The sensor engages in the X-ray unit.
NOTE
When the sensor is being removed, it could be damaged by impact or if
it is dropped.
The sensor contains an integrated vibration sensor to detect impacts or
falls. If the vibration sensor has triggered, guarantee claims become
void.
➢ Do not drop the sensor under any circumstances!
NOTE
Electrostatic charges from persons are discharged on the unit.
This will destroy electrical components in the unit.
➢ Do not touch any electrical components or unprotected plug
contacts.
➢ Discharge yourself by touching a conductive grounded object.