5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.2.4
Adjusting the angle preselection
In the TM1 exposure program, angle (0°, 5°, 10° and 15°) can be preset
for the temporomandibular joint area. This can be helpful if more detailed
analyses of the temporomandibular joint are necessary, but the standard
projections (0°) do not provide an optimal image.
The figure shows the direction in which the slice orientation is being
swiveled during angle preselection.
✔
Level 1 is displayed on the touchscreen.
1. Touch the angle preselection symbol (A) on the right side of the
touchscreen.
ª
A submenu line is opened.
2. Select the corresponding angle preselection.
ª
The selection is highlighted orange in the submenu line. The
selected angle preselection is displayed on the right side of the
touchscreen.
3. Touch the cross on the left side of the submenu line.
or
➢
Touch the angle preselection symbol (A) again.
ª
The submenu line is closed.
ª
The angle preselection is set.
0°
15°
–
0°
15°
–
TM1
A
Ready for exposure
Ready for exposure
Ready for exposure
IMPORTANT
When you confirm the exposure with the R key, the angle setting that
was changed in the submenu line will automatically be reset to the
default setting of 0°.