5.1 Acquiring the X-ray image
11. Align the patient's head according to the Frankfurt horizontal plane
(E).
Tip: The Frankfurt horizontal is used as a reference plane. It runs
between the upper edge of the ear canal and the deepest point of the
lower eye socket edge.
12. Adjust the height of the light localizer using the slider F.
ª
The laser line reflects on the upper edge of the outer ear canal.
13. Correct the patient's head inclination as necessary. Briefly press the
up A and down B height adjustment keys.
ª
The laser line reflects on the lowest point of the lower eye socket
edge.
14. Press the temple support adjustment button I.
ª
When you touch the patient's head, the temple supports stop
automatically.
15. Press the forehead support adjustment key (H).
ª
When you touch the patient's head, the forehead support stops
automatically.
Make sure that the patient's head does not move backwards
when placing the forehead support.
16. Check the patient's position and make any final corrections as
necessary.
17. Swivel the control mirror back into place. Press the right recess on the
touchbar D.
ª
The patient can see himself in the control mirror.
18. Instruct the patient to exhale, place his tongue against the roof of his
mouth, and hold this position until the end of the exposure.
ª
The patient is positioned in the unit.
E
C
B
A
I
H
IMPORTANT
In P1, P2, P10, the slice width is selected automatically for different
dental arches with the temple support setting, and the radiation time is
also changed slightly through this in accordance with the temple support
width which is set.