5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.3.5
Canceling an exposure
An exposure that has been triggered can be canceled again at any time.
✔
The exposure is released.
1. Let go of the exposure release button.
ª
The exposure is immediately terminated.
Message H320 appears.
The exposure data of the terminated exposure is displayed in
Sidexis.
The radiation time and dose-area product flash.
2. Guide the patient out of the unit.
3. Press the R key.
ª
The actual required radiation time is acknowledged.
4. Press the R key again.
ª
The rotating unit moves to the starting position.
ª
The unit is now ready for the next exposure.
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
R
IMPORTANT
The program settings must be checked before the exposure is repeated.
Any changed program settings must be preselected again.