5 Operation
Operating Instructions Orthophos SL 2D / 3D
5.1 Acquiring the X-ray image
5.1.2.4.6
Selecting kV/mA values
85 kV are usually emitted for volume exposures. Preset kV/mA
combinations and radiation times are assigned to the patient symbols,
which are selected according to the patient's size and weight. The
symbols roughly correspond to child, youth/woman, woman/man, hefty
persons. The kV/mA values and radiation times are preset, see also
"Program values for volume exposures".
✔
Level 1 is displayed on the touchscreen.
➢ Touch the required patient symbol.
ª
The selection (A) is highlighted in orange. The selected kV/mA
value is displayed on the right side of the touchscreen and the
radiation time is shown below the program name.
ª
The kV/mA value is selected.
For the USA only: The source to image distance (SID) is displayed on the
touchscreen. It equals 524 mm or 20,63“.
In HD exposure mode: Setting the kV/mA values via the submenu line
If the preset kV/mA combinations do not provide a satisfactory result via
the patient symbols, you can also set the kV/mA values manually in all
programs in HD exposure mode.
✔
Level 1 is displayed on the touchscreen.
1. Touch the kV/mA symbol (B) on the right side of the touchscreen.
ª
A submenu line is opened.
2. Select a kV/mA value. To do this, touch the arrow keys - or +.
Please be aware of the maximum patient dose.
ª
The selected kV/mA value is displayed.
3. Touch the cross on the left side of the submenu line.
or
➢
Touch the kV/mA symbol (B) again.
ª
The submenu line is closed.
ª
The kV/mA value is set.