8 Settings and repair
Dentsply Sirona
8.1 Program values
Operating Instructions Orthophos SL 2D / 3D
8
Settings and repair
8.1
Program values
8.1.1
Panoramic exposure index 1E
Code 1E
This level series is factory programmed for the Federal Republic of
Germany. The code 1E, which specifies a reduced level series for
children and young people, should at least be complied with by law for
new installations or relocation/ shifting of operations since 01/01/1999 in
the Federal Republic of Germany. Furthermore, this level series can also
be applied worldwide.
Level series for code 1E
Program
Pro
-
gram
duration
approx.
Max. ex
-
posure
time
Quickshot
program
duration
approx.
Max.
Quickshot
exposure
time
Factory setting
User-defined values
– Please enter here –
P1
19.0s
12.9s
14.1s
8.0s
14.2s
10.3s
9.0s
5.1s
63/6 63/8 69/12 72/14
P1A
21.8s
15.4s
14.1s
8.0s
18.2s
13.9s
9.0s
5.1s
63/6 63/8 69/12 72/14
P1C
20.1s
13.3s
14.1s
8.0s
17.1s
12.6s
10.5s
5.9s
63/6 63/8 69/12 72/14
P2
16.4s
11.6s
11.5s
6.7s
12.4s
9.4s
7.3s
4.2s
63/6 63/8 69/12 72/14
P2A
18.0s
12.1s
11.5s
6.7s
15.0s
11.8s
7.3s
4.2s
63/6 63/8 69/12 72/14
P2C
16.8s
11.7s
11.5s
6.7s
13.7s
9.7s
8.5s
4.9s
63/6 63/8 69/12 72/14
P10
16.4s
11.6s
11.5s
6.7s
11.4 s
9.4 s
7.3s
4.2s
63/6 63/8 69/12 72/14
P10A
18.0s
12.1s
11.5s
6.7s
15.0s
11.8s
7.3s
4.2s
63/6 63/8 69/12 72/14
P10C
16.8s
11.7s
11.5s
6.7s
13.7s
9.8s
8.5s
4.9s
63/6 63/8 69/12 72/14
P12
11.9 s
4.9s
69/8 75/8 78/14 84/12
BW1
23.0s
23.0s
8.8s
4.5s
63/6 63/8 69/12 72/14
BW2
18.0 s
5.1 s
63/6 66/8 69/12 72/14
TM1.1+TM1.2 16.1+
16.1s
6.4+
6.4s
66/8 69/8 72/14 75/14