Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-33
9146
Ongoing hardware
checks
+15V from pressure solenoid
PCB too low
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9147
Ongoing hardware
checks
+15V from pressure solenoid
PCB too high
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9148
Ongoing hardware
checks
-15V from pressure solenoid
PCB too low
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9149
Ongoing hardware
checks
-15V from pressure solenoid
PCB too high
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9150
Ongoing hardware
checks
+1.2 V reference from pressure
solenoid PCB too low
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9151
Ongoing hardware
checks
+1.2 V reference from pressure
solenoid PCB too high
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9152
Ongoing hardware
checks
+15V from BBU PCB too low
1. Replace BBU PCB.
2. Replace controller PCB.
9153
Ongoing hardware
checks
+15V from BBU PCB too high
1. Replace BBU PCB.
2. Replace controller PCB.
9154
Ongoing hardware
checks
+5V from BBU PCB too low
1. Replace BBU PCB.
2. Replace controller PCB.
9155
Ongoing hardware
checks
+5V from BBU PCB too high
1. Replace BBU PCB.
2. Replace controller PCB.
9156
Ongoing hardware
checks
+5V to breath delivery motor
controller circuits too low
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9157
Ongoing hardware
checks
+5V to breath delivery motor
controller circuits too high
1. Replace pressure solenoid PCB.
2. Replace controller PCB.
9158
Ongoing hardware
checks
+2.5V reference to breath
delivery motor controller
circuits too low
Replace controller PCB.
9159
Ongoing hardware
checks
+2.5V reference to breath
delivery motor controller
circuits too high
Replace controller PCB.
9160
Ongoing hardware
checks
Oxygen regulator pressure
transducer reading too low
1. Calibrate oxygen regulator pressure
transducer.
2. Replace oxygen regulator pressure
transducer.
3. Replace oxygen regulator assembly.
4. Replace pressure solenoid PCB.
9161
Ongoing hardware
checks
Oxygen regulator pressure
transducer reading too high
1. Verify oxygen supply pressure is not too
high.
2. Calibrate oxygen regulator pressure
transducer.
3. Replace oxygen regulator pressure
transducer.
4. Replace oxygen regulator assembly.
5. Replace pressure solenoid PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action