Performance verification
5-10
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
NOTE:
Follow the
PTS 2000 User’s Manual
instructions for setting up the unit with your
computer. Set the PTS 2000
for REMOTE OPERATION.
NOTE:
When using PTS 2000 Breathlab software to measure volume readings, select the
Volume (1 breath)
screen. Use the following settings: Units of measure in
mL
. Mode
ATP
threshold of
10 lpm
. Select multi and activate the arrow button to begin gathering
data on each breath.
Measure PEEP in the LOW PRESSURE
screen. Measure oxygen percentage in the
% Oxygen
screen or the ALL PARAMETERS screen.
2.
Change the settings shown in bold.
3.
Wait until the ventilator delivers at least 10 breaths.
4.
Make the following measurements, and verify the readings are as listed below.
See
15, 16, 17
.
NOTE:
When making PEEP measurements at high rates the readings will fluctuate, making it
difficult to take an accurate reading. To eliminate this difficulty, PEEP can be measured
with the rate turned down to 3 breaths/min.
Control
Setting
HIGH RATE
100 breaths/min
HIGH TIDAL VOLUME
1500 mL
LOW INSP PRESSURE
3 cmH
2
O
LOW TIDAL VOLUME
0 mL
HIGH PRESSURE
90 cmH
2
O
LOW MINUTE VOLUME
0 L
Mode
A/C, VCV (SQUARE flow waveform for software
revision J or later)
RESPIRATORY RATE
30 breaths/min
TIDAL VOLUME
100 mL
PEAK FLOW
30 L/m
PLATEAU 0
s
PEEP/CPAP
3 cmH
2
O
TRIGGER SENSITIVITY
20 L/min
% O
2
21 %