Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-23
7210
POST
Software error
7211
POST
Software error
7212
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7213
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7214
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7215
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7216
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7217
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7218
POST
VENT INOP signal is active
1. See previous diagnostic codes in log for
corrective actions.
2. Run EST.
3. Replace controller PCB.
7219
POST
Operating temperature < 5 °C
Make sure room temperature is not too
low. Make sure that ventilator has warmed
up and exhaled flow attained a steady-
state temperature.
7220
POST
Zeroing of oxygen regulator
pressure transducer failed
Replace oxygen regulator pressure
transducer.
7221-7233
POST
Software error
7234
POST
Controller PCB ID invalid
Replace controller PCB.
7235
POST
Pressure solenoid PCB ID
invalid
Replace pressure solenoid PCB.
7236
POST
BBU PCB ID invalid
Replace BBU PCB.
7237
POST
Loopback test performed on
controller PCB during POST
failed.
Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action