Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-29
9085
Ongoing hardware
checks
Failure to autozero cylinder
pressure transducer
1. Use
Sensor data
function to check P
cyl
reading.
2. Check that autozero solenoid/main
ventilator head harness connection is
secure.
3. Replace pressure solenoid PCB.
9086
Ongoing hardware
checks
Inspiration pressure out of
bounds
1. Use
Sensor data
function to check P
i
reading.
2. Replace pressure solenoid PCB.
9087
Ongoing hardware
checks
Exhalation pressure out of
bounds
1. Use
Sensor data
function to check P
e
reading.
2. Replace pressure solenoid PCB.
9088
Ongoing hardware
checks
Exhaled flow out of bounds
1. Make sure room temperature is within
specifications. Make sure that
ventilator has warmed up and exhaled
flow attained a steady-state
temperature.
2. Replace flow sensor.
3. Replace pressure solenoid PCB.
9089
Ongoing hardware
checks
Oxygen regulator pressure
transducer reading out of
range
1. Make sure oxygen source pressure is at
least 40 psi.
2. Calibrate oxygen regulator pressure
transducer.
3. Check for proper installation of oxygen
regulator pressure transducer and
connection of sensor to pressure
solenoid PCB.
4. Replace oxygen regulator pressure
transducer.
5. Replace oxygen regulator assembly.
6. Replace pressure solenoid PCB.
9090
Ongoing hardware
checks
Motor controller chipset
failure
Replace controller PCB.
9091
Ongoing hardware
checks
Motor controller chipset
failure
Replace controller PCB.
9092
Ongoing hardware
checks
Piston motion error
1. Perform gear mesh procedure, as
required.
2. If piston scrapes, replace piston/cylinder
assembly.
3. Replace motor/encoder.
4. Replace BBU PCB.
5. Replace controller PCB.
9093
Ongoing hardware
checks
Unexpected interrupt to motor
controller chipset
Replace controller PCB.
9094
Ongoing hardware
checks
Unexpected interrupt to motor
controller chipset
Replace controller PCB.
9095
Ongoing hardware
checks
Motor fault reported to motor
controller chipset
1. Replace piston/cylinder assembly.
2. Replace BBU PCB.
3. Replace motor/encoder.
4. Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action