Service and utility functions
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
4-7
Table 4-6: UI/HW sensors
Specific parameter
Description
Should read within this range...
UI M24V
+24 V on UI PCB (in V)
19.5 to 30.7 V
UI M5VD
+5 V on UI PCB (in V)
4.7 to 5.4 V
Cara 1/2 V
Ventilator model
740: approx. 0
760: 1023 counts
Cara ID
Unused.
CB HW ID
Controller PCB hardware revision
Depends on PCB revision
MB HW ID
BBU PCB hardware revision
Depends on PCB revision
PS HW ID
Pressure solenoid PCB hardware revision
Depends on PCB revision
OP HW ID
Reserved for future use
For future use
Table 4-7: Pressure sol sensors
Specific parameter
Description
Should read within this range...
P15V
+15 V measured on pressure solenoid PCB
(in V)
14.0 to 15.9 V
N15V
-15 V measured on pressure solenoid PCB
(in V)
13.7 to 15.7 V
1V2 REF 1
+1.2 V reference 1 on pressure solenoid
PCB (in V)
1.2 to 1.3 V
1V2 REF 2
+1.2 V reference 2 on pressure solenoid
PCB (in V)
1.2 to 1.3 V
Table 4-8: Other data
Specific parameter
Description
Should read within this range...
Peep I*
PEEP pump current (in counts)
183 to 818 counts
% O
2
*
Monitored oxygen percentage
9 to 109%
Delivered volume*
Delivered volume (in mL/min)
40 to 2000 mL
Humidifier type
As specified during SST/EST
HME, Dual heated wire, No heated wire
Tubing compliance
Ventilator breathing circuit compliance
determined during the last run of SST/EST
(in mL/cmH
2
O)
Pediatric or Adult
ET size
As specified during SST/EST
3 to 9 mm
Operational hours
Total hours ventilator has operated
*Varies with patient settings during ventilation. Can also be read when not ventilating.