Theory of operation
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
2-5
Table 2-1 provides some pertinent specifications and other data for the ventilator’s
pneumatic components.
Table 2-1: Pneumatic component data
Component
Data
Check valve, air intake,
inlet
Opens to draw in room air. Closes during oxygen delivery to reduce
noise.
Check valve, air intake
pressure relief
Remains closed during normal operation. Opens at or below
5 cmH
2
O to prevent excess pressure in the intake duct.
Check valve, cylinder inlet
Opens during piston retraction to draw in room air and/or oxygen
Check valve, cylinder outlet
Opens to deliver mixed gas to the patient
Check valve, exhalation
Opens during exhalation to permit exhaled gas into the exhalation
system. Prevents rebreathing.
Collector vial
Collects moisture in the exhaled gas to keep water from occluding
the expiratory filter or entering the exhalation system.
Exhalation valve
Piloted by exhalation solenoid. Opened during exhalation (as
required to maintain PEEP/CPAP). Closed during inspiration.
Filter, air intake
≥ 5
μ
m retention
Filter, expiratory
> 0.3
μ
m nominal (and larger) retention at flow of 100 L/min
Filter, inspiratory (main
flow)
> 0.3
μ
m nominal (and larger) retention at flow of 100 L/min
Filter, PEEP reservoir
> 0.3
μ
m nominal (and larger) retention at a flow of 10 L/min
Fitting, oxygen (source)
DISS (diameter index system standard) male. DISS female and male,
NIST (non-interchangeable screw thread), Air Liquide™*, Australian
type, and Dräger™* oxygen hose kits available, using adapters as
necessary.
Heater assembly,
exhalation
Two 10 W heaters that maintain temperature of exhalation assembly
walls above condensation point
Piston/cylinder assembly
(includes piston/cylinder,
motor/encoder, and
optoswitches)
Includes a brushless DC motor. Can deliver up to 2 L nominal.
Clearance of 50
μ
m (0.002 in.) between piston and cylinder wall.
Pump, PEEP
Supplies PEEP pilot pressure of 0 to 25 cmH
2
O
Regulator, oxygen
Output of 33
±
5 psi (227.46
±
34.48 kPa) throughout supply pressure
range of 40 to 90 psi (275 to 620 kPa) (no flow). Includes internal
filter (5
μ
m retention) and outlet filter (> 30
μ
m retention). Maximum
3 L/min bleed.
Sensor, exhalation flow
A mesh screen with pressure taps on both sides. Pressure drop across
screen monitored by a differential pressure transducer and used to
determine flow.
Sensor, oxygen
Measures percentage of oxygen in inspired gas (based on partial
pressure of oxygen). Range: 18 to 103% oxygen.