General information
1-16
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Flat-bladed screwdriver
Local supplier
Various service procedures
Flat-bladed screwdriver with long
(>20-cm) shank
Local supplier
Piston/cylinder assembly retaining
screws
Phillips screwdriver, no. 1
Local supplier
Cart
NOTE:
Use Phillips screwdrivers only to remove Phillips-head screws. Using POZIDRIV™* screwdrivers on Phillips heads
may damage the screw heads.
Drill with 1/8-in. (3-mm) bit
Local supplier
Removing serial number plate
Electrical safety analyzer, capable of
measuring ground resistance and
leakage current
Fluke ESA620 or equivalent
Fluke Model ESA 620 or
equivalent
http://www.flukebiomedical.com
6920 Seaway Blvd.
Everett, WA USA 98203
Telephone: (425) 446-6945
Toll Free (US): (800) 850-4608
Electrical safety test
Digital multimeter (DMM) accurate to
3 decimal places
Local supplier
Performance verification
Multimeter patch cord set, 0.025 square
receptacle (Includes red cord
(P/N G-061579-00) and black cord
(P/N G-061580-00))*
G-061567-00 or equivalent
Performance verification
Pneumatic calibration analyzer or
equivalent devices capable of
measuring oxygen percent, flow, BTPS
volume, pressure, and barometric
pressure. Oxygen analyzer connector
tee. Required accuracies:
•
Flow: 2.75% of reading ±0.05 slpm
•
Volume: 2% of reading or ±1 digit
•
Low pressure (-150 to +150 cmH
2
O):
0.75% of reading ±0.04 cmH
2
O
•
High pressure (0 to 150 psig):
1.0% of reading ±0.1 psi
•
Oxygen percentage: ±2% oxygen
Puritan Bennett™ PTS 2000
Performance Test System
(4-074686-00)
Performance verification, EST,
oxygen regulator pressure
transducer calibration
ESD-safe vacuum cleaner with 0.2
μ
m
filter (rated for photocopiers and laser
printers)
Local supplier
General cleaning of ventilator
interior
Static-dissipative field service kit
(includes wrist strap, static-dissipative
mat, and ground cord)
*
G-061661-00 or equivalent
Various service procedures
Oxygen source, 40 to 90 psi (275 to
620 kPa); (≥ 50 psi (345 kPa) required
for oxygen sensor calibration check)
Local supplier
EST, performance verification,
oxygen sensor calibration check
Table 1-7: Tools, equipment, and service materials (continued)
Description
Manufacturer and model or
Covidien part number
Where used
Summary of Contents for Puritan Bennett 700 Series
Page 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...