Diagnostic codes
6-40
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
13007
Ventilator status
Ventilator reset occurred while
waiting for user to confirm
settings
13008
Ventilator status
SST passed
None required.
13009
Ventilator status
SST incomplete
None required.
13010
Ventilator status
SST fault
None required.
13011
Ventilator status
SST failure
None required.
13012
Ventilator status
EST passed
None required.
13013
Ventilator status
EST incomplete
None required.
13014
Ventilator status
EST fault
None required.
13015
Ventilator status
EST failure
None required.
13016
Ventilator status
Ventilator reset occurred in
normal ventilation mode
13017
Ventilator status
Ventilator reset occurred while
waiting for user to confirm
settings
13018
Ventilator status
SST failed and ventilator
restarted with MENU key
pressed down, causing the
ventilator to enter VENT INOP
None required.
13019
Ventilator status
(technical alert --
ABNORMAL RESTART)
A field in the setting area of
NVRAM is corrupted (for
example, bad NVRAM data
downloaded to ventilator).
The ventilator resets, returns settings to
default values, and stores the error in
NVRAM. Check and confirm settings,
including alarms.
14000
Communications
failure
An invalid request was
received on Port A or Port B
1. Check that external communications
device is valid for use with the 700
Series Ventilator.
2. Check all cable connections between
the Communications panel,
Communications option assembly, and
controller PCB.
3. Replace Communications option
assembly.
14001-
14xxx
Communications
failure
Software error
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action
Summary of Contents for Puritan Bennett 700 Series
Page 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Page 439: ......