Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-31
9113
Ongoing hardware
checks
Real-time clock failure
Replace controller PCB.
9114
Ongoing hardware
checks
Real-time clock failure
Replace controller PCB.
9115-9124
Ongoing hardware
checks
Bad NVRAM data or NVRAM
1. See Section 6.2.
2. Restore NVRAM data.
3. Replace NVRAM.
9125
Ongoing hardware
checks
Exhalation flow sensor
pressure transducer failure
1. Make sure room temperature is within
specifications. Make sure that
ventilator has warmed up and T
xdcr
attained a steady state.
2. Replace pressure solenoid PCB.
9126
Ongoing hardware
checks
Mismatch between BD and UI
microprocessors’ inspiration
pressure readings
1. Verify sampling tubes are connected to
correct pressure solenoid PCB
connectors.
2. Replace pressure solenoid PCB.
3. Replace controller PCB.
9127
Ongoing hardware
checks
Mismatch between BD and UI
microprocessors’ exhalation
pressure readings
1. Verify sampling tubes are connected to
correct pressure solenoid PCB
connectors.
2. Replace pressure solenoid PCB.
3. Replace controller PCB.
9128
Ongoing hardware
checks
Not Ventilating signal used in
UI/BD communications is not in
expected state.
Replace controller PCB.
9129
Ongoing hardware
checks
Not Ventilating signal used in
UI/BD communications is not in
expected state.
Replace controller PCB.
9130
Ongoing hardware
checks
Exhaled flow out of bounds
1. Make sure room temperature is within
specifications. Make sure that
ventilator has warmed up and exhaled
flow attained a steady-state
temperature.
2. Replace flow sensor.
3. Replace pressure solenoid PCB.
9131
Ongoing hardware
checks
Cylinder pressure reading out
of range
1. Make sure inspiratory filter outlet is not
blocked and air intake filter is not
occluded.
2. Replace pressure solenoid PCB.
9132
Ongoing hardware
checks
Flow sensor pressure
transducer reading too low
1. Make sure room temperature is not too
low. Make sure that ventilator has
warmed up and T
xdcr
attained a steady
state.
2. Verify tubes between flow sensor and
pressure transducer are securely
connected.
3. Make sure EXHAUST port is not
blocked.
4. Replace pressure solenoid PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action