Performance verification
5-14
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
2.
Wait three minutes for oxygen analyzer reading to stabilize. Verify oxygen analyzer
reads between 27 and 33% and that no oxygen alarms activate (if applicable).
See
Section 5.7 18.
3.
Change the ventilator settings shown in
bold
.
4.
Wait 1.5 minutes for oxygen analyzer reading to stabilize. Verify oxygen analyzer reads
between 57% and 63%, and that no oxygen alarms activate (if applicable).
See
Section 5.7 18.
5.
Change the ventilator settings shown in
bold
.
Control
Setting
HIGH RATE
100 breaths/min
HIGH TIDAL VOLUME
1500 mL
LOW INSP PRESSURE
3 cmH
2
O
LOW TIDAL VOLUME
5 mL
HIGH PRESSURE
90 cmH
2
O
LOW MINUTE VOLUME
0 L
Mode
A/C, VCV (SQUARE flow waveform for software
revision J or later)
RESPIRATORY RATE
20 breaths/min
TIDAL VOLUME
600 mL
PEAK FLOW
60 L/min
PLATEAU
0 s
PEEP/CPAP
0 cmH
2
O
TRIGGER SENSITIVITY
20 L/min
% O
2
60
Control
Setting
HIGH RATE
100 breaths/min
HIGH TIDAL VOLUME
1500 mL
LOW INSP PRESSURE
3 cmH
2
O
LOW TIDAL VOLUME
5 mL
HIGH PRESSURE
90 cmH
2
O
LOW MINUTE VOLUME
0 L
Mode
A/C, VCV (SQUARE flow waveform for software
revision J or later)
RESPIRATORY RATE
10 breaths/min
TIDAL VOLUME
1000 mL
PEAK FLOW
80 L/min
PLATEAU
0 s