Diagnostic codes
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
6-9
6028
BIOT (technical alert --
CONTACT SERVICE)
Safety valve should be closed,
but measured current is too
high
1. Check that safety valve/ventilator head
harness connection is secure.
2. Check safety valve resistance (10.5 to
14
Ω
).
3. Replace safety valve.
4. Replace pressure solenoid PCB.
6029
BIOT (technical alert --
CONTACT SERVICE)
No current to ALARM light
when it should be on
1. Replace UI display PCB.
2. Replace controller PCB.
6030
BIOT (technical alert --
CONTACT SERVICE)
There is current to ALARM
light when it should be off
1. Replace UI display PCB.
2. Replace controller PCB.
6031
BIOT (technical alert --
CONTACT SERVICE)
No current to CAUTION light
when it should be on
1. Replace UI display PCB.
2. Replace controller PCB.
6032
BIOT (technical alert --
CONTACT SERVICE)
There is current to CAUTION
light when it should be off
1. Replace UI display PCB.
2. Replace controller PCB.
6033
BIOT (technical alert --
CONTACT SERVICE)
Battery current low
1. Replace appropriate battery.
2. Replace BBU PCB.
3. Replace controller PCB.
6034
BIOT (technical alert --
CONTACT SERVICE)
Battery current high
1. Replace appropriate battery.
2. Replace BBU PCB.
3. Replace controller PCB.
6035
BIOT (technical alert --
CONTACT SERVICE)
PEEP pump current low
1. Verify resistance of PEEP pump is
between 12 and 15
Ω
measured when
pump is at ambient temperature. If it is
not replace pump.
2. Check continuity of wiring to PEEP
pump. Replace harness if necessary.
3. Replace pressure solenoid PCB.
4. Replace BBU PCB.
5. Replace controller PCB.
6036
BIOT (technical alert --
CONTACT SERVICE)
PEEP pump current high
1. Verify resistance of PEEP pump is
between 12 and 15
Ω
measured when
pump is at ambient temperature. If it is
not replace pump.
2. Check continuity of wiring to PEEP
pump. Replace harness if necessary.
3. Replace pressure solenoid PCB.
4. Replace BBU PCB.
5. Replace controller PCB.
6037
BIOT (technical alert --
AIR INTAKE ABSENT)
Air intake filter switch open
1. Verify air intake filter is installed and
that it engages air intake filter switch.
2. Check that air intake filter switch/
ventilator head harness connection is
secure.
3. Replace air intake filter.
4. Replace air intake filter switch.
5. Replace pressure solenoid PCB.
6038
BIOT (technical alert --
CONTACT SERVICE)
NVRAM failure
1. Replace NVRAM.
2. Replace controller PCB.
Table 6-1: Diagnostic codes (continued)
Code
Software
Possible cause
Corrective action