Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-33
Figure 8-27. Removing oxygen regulator/oxygen solenoid assembly
5.
To remove the grounding strap from the regulator, use a 3-mm hex key to loosen the
two M3 x 25 screws on the grounding strap, then remove the grounding strap from the
regulator.
6.
To separate the oxygen solenoid assembly from the oxygen regulator, use a 2.5-mm allen
key to remove the four hex screws (Figure 8-28).
Figure 8-28. Separating the oxygen solenoid assembly from the oxygen regulator
7-00263
Oxygen
solenoid
assembly
Oxygen
regulator
Mixing
manifold
7-00196