Service and utility functions
4-2
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
4.1.2 Summary of functions
Table 4-1 and Table 4-2 summarize the ventilator’s
Service
and
Utilities
functions.
4.1.3 Navigating the
Service
and
Utilities
menus
Figure 4-1 shows you how to navigate through all the
Service
and
Utilities
menu functions. In
general, the ventilator controls serve these functions:
• The knob lets you locate a function or menu.
• The ACCEPT key selects the function or allows you to proceed to the function’s next
prompt.
• The CLEAR key returns you to the start of the function or to the previous menu.
Table 4-1: Service functions
Description
Purpose
Sensor data
Primarily for factory use. During ventilation displays monitored parameters,
including pressures, temperatures, voltages, battery current, PEEP motor on-time,
delivered volume, flow BTPS, and oxygen percentage.
EST
Runs extended self-test (see Chapter 3 for details on EST).
Diagnostics/Calib
Enables the serial port and up/download (if the Communications option is installed,
you can configure serial ports A and B), runs individual EST tests, performs
calibrations, erases the contents of the test and alert logs, resets service data, and
lets you review and manually update calibration constants.
Review alert log
Lets you read the alert log contents
Test data
Lets you read the test log contents and the EST test results
Table 4-2: Utilities menu functions
Description
Purpose
Pressure format
Lets you choose the unit for display of pressures (cmH
2
O or hPa).
Date & time format
Lets you choose the format for display of date (European--with day first, followed
by month, or US--with month first, followed by day) and time (12- or 24-hour
clock).