General information
1-18
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
1.9 Periodic maintenance
Table 1-8 lists the periodic maintenance activities (other than patient system maintenance)
required for the 700 Series
Ventilator. The hours remaining until service is due are displayed
when POST is run and through the
Service summary
menu option (see the
Operator’s
Manual
). For patient system maintenance, consult the
Operator’s Manual
.
Table 1-8: Schedule of periodic maintenance
Interval
Part
Activity
250 hours or 1 month of use (or
more often, if required)
Main fan filter (G-060531-00)
Either vacuum filter or wash
filter in a warm detergent
solution, rinse, and dry well.
Replace filter when it shows signs
of wear or when a FAN FAILED
ALERT alarm occurs.
1000 hours or 3 months of use (or
more often, if required)
Air intake filter (G-060457-00)
Replace at recommended
interval or when an AIR INTAKE
BLOCKED alarm occurs.
Every year or per your hospital’s
protocol
Entire ventilator
Electrical safety test
Every quarter
Backup alarm function
Test backup alarm function per
Section 8.10.3.2 steps
Every two years
Entire ventilator
Performance Verification
15,000 hours of operation
Entire ventilator
Install 15,000-hour preventive
maintenance kit
(P/N G-061166-00), including
performance verification.
30,000 hours of operation
Entire ventilator
Install 30,000-hour preventive
maintenance kit
(P/N G-061167-00), including
performance verification.
Every 2 years or as necessary
•
Internal battery
(G-061139-00)
•
External battery
(G-061140-00)
•
Oxygen sensor (G-062010-00
for sensor only; G-062009-00
if replacing a sensor with
integral harness)
Replace.
NOTE:
Using ventilator menu functions, you can view the remaining
estimated life of the internal and external batteries, in hours.